indicates the typical heat dissipation at nominal load and specify cooling air flow in two
situations:
•
1000 m (3280 ft) altitude, cooling air between 25°C (77°F) and 30°C(86°F)
•
250 m (820 ft) altitude, cooling air below 18°C (64°F)
Table 3-4:
Fluoro UPS
Typical heat output
On–line cooling air flow at 30°C /
86°F
Typical on-line cooling air flow at
18°C / 64°F
CE version
1.50 kW (5101 BTU/hr)
625 m3/h (368 CFM)
227 m3/h (134 CFM)
UL version
4720 BTU/hr (1.38 kW)
301 CFM (511 m3/h)
119 CFM (203 m3/h)
2.2.3 IVUS Option
2.2.3.1 Environmental Limitations in Use (room climate)
Table 3-5: IVUS Rev 1
Humidity (non condensing) :
30 - 80 %
Temperature (at 20% Relative Humidity ) :
15 - 32 C
(60 - 90 F)
Altitude (Relative to see level) :
max 3048 m
(max 10000 ft)
Table 3-6: IVUS Rev 2
Humidity (non condensing) :
30 - 80 %
Temperature (at 20% Relative Humidity ) :
5 - 35 C
(41 - 95 F)
Altitude (Relative to see level) :
max 3048 m
(max 10000 ft)
2.2.3.2 Heat Output
Device
Location
Heat output
PC
Control room
335 W – 1206 BTU/h
Monitor
Control room
125 W – 427 BTU/h
Printer
Control room
67 W – 229 BTU/h (idle)
CAUTION
Take any steps necessary to provide correct temperature and dust free
environment for correct operation of PC in the control room.
2.3 IEC60601–1–2 Electromagnetic Standard Compliance & Documentation
2.3.1 General Scope
This equipment complies with IEC-60601-1-2: Edition 2.1 and Edition 3 EMC standard for medical
devices.
The Innova system is suitable to be used in the electromagnetic environment, as per the limits &
recommendations described in the tables here after:
•
Emission Compliance level & limits (
).
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 3 Physical Requirements of room
59