Illustration 6-2: Power distribution - UL
Cable
CE UPS configuration
UL UPS configuration
Max length
Recommended (1)
Type
Max gauge
Recommended (1)
Type
R rectifier
5 x 10 mm²
3ph+N+PE
AWG3
AWG6
3ph+N+PE
24 m (78ft 8in)
L out /load
5 x 10 mm²
3ph+N+PE
AWG3
AWG6
3ph+N+PE
24 m (78ft 8in)
(1) Size and type may be adapted locally per local regulation.