NOTE:
You should consider running the ECG cable through the conduit in case, the
customer requires interconnecting to a NON Mac-Lab ECG at the table base.
In this case you shall hide the unconnected cable in the conduit for future
expansion options.
NOTE:
Same applies to the 2nd Ethernet cable, if a later extension of the IVUS remote
controls expected.
•
Ensure sufficient cable length on the Procedures Room end of the cables according to the
table below
Cable
Free length from the conduit exit
IVUS PIM cable *
2.4 m (8 ’)
FFR cables for Pimette **
0m (0’)
Control Station cable
2.1 m (7 ’)
Joystick cable
2.7 m (9 ’)
ECG cable
Depends on the ECG location
* Consider the 12’ length of the PIMr cable, if PIMr exists.
** Pimette has 3 m cable.
NOTE:
The incorrect cable length might cause reliability issues during the operation !
Make sure, that the out of conduit parts of the cables with the connectors are properly protected
during the rest of the installation
6.4 Electrical Requirements
The power for the IVUS hardware components delivered with the product (PC, monitor, Video
Switch and printer) is supplied through the IVUS Safety Isolation Transformer.
The customer is in charge to provide a wall outlet to supply the Safety Isolation Transformer. This
outlet shall be located in the control room suitable to the IVUS components positioning. The
maximum rated power of the Safety Isolation Transformer is 600 VA, the maximum continuous
system power consumption for s5i and s5iGE IVUS systems is 400 VA.
The s5i and s5iGE IVUS systems can be configured to work on 100, 120, or 230 VAC. The
difference between the supported configurations is limited switching the fuse values. See table
below :
Line Voltage
Fuse value
100 VAC
8 A
120 VAC
6.3 A
230 VAC
3.15 A
Frequency: 50/60 Hertz
Innova 2121-IQ/3131-IQ Biplane Cardiovascular Imaging System Pre-Installation Manual
GE Healthcare
Direction 5177951-4-1EN, Revision 1
Chapter 6 Electrical Connections
165