Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
97
2
ROOM LAYOUT DRAWINGS
See Illustration 45, Illustration 46, Illustration 47, Illustration 48 and Illustration 49, for recommended
room layouts for an Innova LC system.
3
ROOM SPEAKER
If the Advantx console tone cannot be heard from the exam room, a remote loud–speaker must be
installed in that room; the loud–speaker, which must be purchased locally, has to be compliant with
the local legal requirements (if any; e.g. CE Marking for EEC).
The loud–speaker is to be connected to the console (1/4”, serial 3 conductor connector).
In US, a Radio Shack loud–speaker (catalog #40–1383), tested and validated, can be used.
4
ROOM LIGHTING
See Illustration 49, for a recommended room lighting layout for an Innova LC system.
4-1
Requirements for lighting
Requirement for lighting concern the following, general, light–technique characteristics:
Illuminator level.
Lighting distribution.
Preventing the operator from being dazzled by the light (by direct light sources or by reflection on
bright objects).
The Illumination level must be compliant with established lighting technical rules and be as constant
as possible.
5
EMERGENCY
During an examination, any operator can encounter two main cases of failures.
5-1
Main power supply cut
In this case, refer to Chapter 6, section 11.
5-2
System itself fall into failure
When the system fall into failure with a patient on table during an examination, the operator can
require a help with a Surgical Imaging mobile unit to finish the examination.
In this case a wall outlet single phase + ground is required to feed the mobile. It is also requires a free
space around the patient table to proceed with the mobile instead of Innova LC Positioner. The table
has to rotate to 90
.
Refer to Illustration 50 and Illustration 51 for a surgical imaging mobile use.