Innova 2000 Cardiovascular Imaging System
GE Medical Systems
Pre–Installation Manual
REV 5
pim 2337741–100
118
7
INJECTORS
7-1
Remote Injector (rack mount)
A power cord is supplied with the injector.
If the voltage is 110 Vac, the cord should be connected inside the bulkhead positioner cabinet, but it is
highly recommended to connect it to a wall outlet fed by the main disconnect room device.
7-2
Pedestal Injector
A power cord is supplied with the injector.
In all cases (110 Vac or 230 Vac), it will be connected to a wall outlet near the operator location, fed by
the main disconnect room device.
7-3
Injector L.F. ILLUMINA
As a result of power demand availability for all functions of this new injector, in any cases it will be
connected to a wall outlet sized to 1500 W, 110 VAC or 230 VAC as nominal voltage. The wall outlet is
fed by the main disconnect room device.