HygroPro User’s Manual
Page 37
Chapter 3: Service & Maintenance
3.3.1 Preparing to Clean the Probe (cont.)
To clean the moisture probe, the following items are required:
•
Two glass (
NOT
metal) containers with approximately 300 ml of
reagent-grade hexane or toluene in each container.
•
One glass (
NOT
metal) container with approximately 300 ml of
distilled (
NOT
deionized) water.
IMPORTANT:
Make sure the containers are deep enough to completely
submerge the moisture probe. Do not place the
transmitter module into the solvents. Insert only the
moisture sensor into the solvents.
•
Rubber or latex gloves
•
An oven set at 50°C ± 2°C (122°F ± 3.6°F)
•
1-1/8” wrench
3.3.2 Replacing the RTE
To maximize
HygroPro
performance, GE recommends recalibration
of the aluminum oxide moisture sensor on the
Replaceable
Transducer Element
(
RTE
) every 6 to 12 months. The optimum
interval depends on the specific application. To accomplish this,
either return the RTE to GE for recalibration or install a new RTE.
The
HygroPro
electronics will automatically read and store the
calibration data whenever a new or recalibrated RTE is installed.
IMPORTANT:
The programmed probe calibration data should
not
be
changed without consulting GE.
Summary of Contents for HygroPro
Page 1: ...GE Measurement Control Moisture 916 099 Rev D November 2012 HygroPro User s Manual ...
Page 2: ......
Page 4: ...ii no content intended for this page proceed to next page ...
Page 24: ...Chapter 1 Installation Page 18 HygroPro User s Manual ...
Page 40: ...Chapter 2 Operation Page 34 HygroPro User s Manual ...
Page 48: ...Chapter 3 Service Maintenance Page 42 HygroPro User s Manual ...
Page 58: ......
Page 59: ......