41
5
Transmembrane Pressure (TMP) Test
Following a successful integrity test, with the system still
flushing through a tap, the TMP must again be measured and
recorded on the service tag. To perform this test, follow the
steps in the Transmembrane Pressure Test section.
6
Battery Replacement
The two AA-sized batteries in the system controller should
be replaced once a year. The batteries are located under
the front cover. Ensure that the settings of the controller are
still appropriate. Refer to Installing the Controller and the
Controller Settings sections.
7
Maintenance Service Tag
Ensure the maintenance service tag has been completed.
8
Final Flush by Homeowner
Have the homeowner continue to run a cold water tap for 60
minutes. This is required to meet NSF drinking water standards
and to remove residual chlorine. The homeowner may notice
chlorine taste/smell when a tap is first opened during a day or
two following the sanitization procedure.
Annual Maintenance Check Summary
•
If external prefilter is used, the external prefilter cartridge
is replaced
•
“Pre-cleaning” TMP completed
•
“Pre-cleaning” TMP recorded on service tag
•
System depressurized and opened
•
Backpulse tank pressure checked
•
System soaked with sanitizer (either bleach or MC1) for
60 minutes
•
System flushed for 10 minutes
•
Integrity test completed
•
Either carbon prefilter replaced or stainless steel prefilter
cleaned and reused
•
“Post-cleaning” TMP completed
•
“Post-cleaning” TMP recorded on service tag
•
Batteries (2 AA) are replaced
•
Service tag completed
•
Homeowner informed to leave cold tap on for
60 minutes
Cleaning the Stainless Steel Facade or Nameplate
If required, the surface may be cleaned using a mild soap
solution. Do not use any other cleaners or abrasive materials
on the stainless steel surface as these cleaners may damage
the labels. Wipe surface dry with a soft cloth afterward.