GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-84
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-37
Cup Holder
8-2-37-1
Tools
•
No need tool
8-2-37-2
Removal procedure
1.) Remove the cup holder.
2.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-37-3
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-125 Cup holder
Table 8-38 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5394152, Rev 1+, Section 8-2-37. Equipment
passed all required tests and is ready for use. “