GE H
EALTHCARE
D
IRECTION
5394152,
Revision 5
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Page 8-10
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-3
17 INCH LCDMON ASSY
8-2-3-1
Tools
•
L wrench
8-2-3-2
Preparation
•
Shut down the system and switch off the main breaker
8-2-3-3
17" LCDMON ASSY Replace Procedure
1.) Remove the LCD Cable cover.
2.) Unscrew 1 screw (2306565,BH M4 x 16WHT) to remove the power code bracket assembled under
the LCD cable cover.
3.) Disconnect power and DVI cable.
Figure 8-14 Removing the LCD Cable Cover
Figure 8-15 Removing the power code bracket