Operating Manual ERESCO MF4 Type Series
Copyright by GE Sensing & Inspection Technologies GmbH Bogenstrasse 41 D - 22926 Ahrensburg
The warm-up program will be started by pressing the button X-RAY ON
.
If the equipment will be switched off e.g. by uneven operation of the X-ray tube or by a fault
during the warm-up program, a corresponding message will appear on the display.
The procedure in case of a shut-down by a fault or an uneven operation of the X-ray tube,
is carried out as discribed in section 2.2.4, page 21 and 22.
After this type of technical fault (or X-RAY OFF) during the extended warm-up mode, the
user still has
four
further attempts to warm-up the tube. After this, the warm-up will be
stopped and can only be restarted after a renewed mains-ON.
If the warm-up program is terminated five times, the following message appears in the
comment line on the display
:
The unit can only be restarted after a renewed mains ON.
In this case, a service technician should be consulted.