K319 Issue 2
EN-47
Englis
h
Special functions
1.
Steps;
10 programmable, 10%, 20%, 25% div. step-
ping by key or adj. timer
2.
Ramp;
programmable travel time for up/down and
dwell
3.
Scaling;
in 5 digits and sign on all electrical ranges
4.
Temp. XMT cal.;
both measure and source readings
in temp. units. Cal.function extended for all output
functions
5.
Temp. XMT sim.;
mA source reads in temperature
units
6.
Signal converter;
any measure into any source,
fully isolated
Note:
Thermocouple accuracies do not include cold junction
compensation errors
Reference
22 °C 1, RH 45% ±15
Accuracies
1 year for 17 °C to 27 °C; outside these
limits 0.0005% / °C typical on mV
Cold Junction error
±0.4 °C (0.8 °F) max. in transient
+0.2 °C (0.4 °F) max. when steady
Resolution error
Included in range error
Long term drift after 1 year
0.005% / year or 1000 hours operating
Cold Junction modes
Internal, Manual, External
Temperature scale
Selectable IPTS 68 or ITS 90
Operating temp.
-10 °C to 50 °C (14 °F to 122 °F)
Storage temp.
-20 °C to 70 °C (- 4 °F to 158 °F)
Relative humidity
0 - 90% non-condensing
Electrical protection
50 V, 30 sec. max.
Ohms source to mV-ground terminal; 5 V max.
Readings
Multilingual. Temp. °C or °F
Batteries (LR14 or C)
4x 1.5 V alkaline or 1.2 V NiCd
Battery life
Alkaline; ;10 hours at 20 °C (68 °F)
Ni-Cd;
;8 hours at 20 °C (68 °F)
Battery life 12 mA sourcing
Alkaline;
5.5 hours at 20 °C (68 °F)
12 mA sourcing
Ni-Cd;
5 hours at 20 °C (68 °F)
Low battery warning
Shows a battery symbol
Line power
With optional adaptor 115 - or 230 VAC
Recharge NiCd'
With optional adaptor 115 - or 230 VAC
User terminals
Gold plated, standard 4 mm plugs
T/C comp.wire terminals
Screw fixable, 2 mm diameter max.
Protection IP
53
Size
210 x 120 x 50 mm without case
(8.4 x 4.8 x 2 inch)
Carrying case
Cordura
Weight
1.2 kg (2.7 lbs) with batteries
Keystroking
9 operation sequences storage
Contact change detect.
Reading freezes on open or close
Switch resistance
Switch resistance measurement
during switch test.
PCMCIA station
PCMCIA card type 1 or 2
Detect. level freq.in
Adjustable; 10 mV to 5.0 V
Pulse source level
Adjustable; 0 - 24 V, 0.1 V resolution,
2% accuracy
Pulse source form
Symmetrical square, zero based - 70 mV
Pulse source speed
Adjustable in Hz or Baud rate
LCD display
Full graphic, with contrast control and
backlight timer
Ohms simulation
0.18 to 5.0 mA excitation current either polarity
Ohms input terminals
2, 3 or 4 wire automatic mode select.
True 4 - wire system
Specifications subject to change without notice
Summary of Contents for Druck TRX-II
Page 2: ......
Page 6: ...EN iv K319 Issue 2 ...
Page 7: ...K319 Issue 2 EN v English ...
Page 8: ...EN vi K319 Issue 2 ...
Page 9: ...K319 Issue 2 EN vii English ...