EN-14
K319 Issue 2
Ohms
Select OHMS and DIRECT in the measure menu and press EXE to
open the working screen. The upper window in the working screen
indicates the measured ohms reading. The range is 0-2000 Ohms
split into two (auto) ranges of 0 to 400 Ohms and 400 to 2000 Ohms.
The excitation current is 0.9 mA.
The wiring can be 2, 3, or 4 wire. The 4 wire method provides a true
measurement of voltage drop across unknown resistance at a
constant excitation current.
Frequency
Select FREQ in the measure menu and press EXE. Select FREQ
and DIRECT in the freq. menu and enter a trigger level between 10
mV and 5V. Press EXE to open the working screen.
The upper window of the working screen indicates the measured
frequency reading. The range is 0-20.000 Hz split into three
(auto)ranges of 0-655 Hz, 655-1310 Hz and 1310-20.000 Hz.
Press # and
simultaneously to increase the trigger level.
Press # and
simultaneously to decrease the trigger level.
Read section "Scaled Readings" to use the SCALE option.