IMPORTANT NOTE
Unless otherwise indicated in the text, the operating instructions
contained in this publication apply to both the TRXII Multi-Calibrator
instruments and the TRXII IS Multi-Calibrator instruments.
Only the intrinsically safe version of this Multi-Calibrator instrument
(TRX II IS) may be used in hazardous areas and the following
general warnings and conditions of use summarized below are
applicable. Pages 36 to 38 provide supplementary information for
the intrinsically safe variants
Before using TRXII IS, intrinsically safe versions of this Multi-
Calibrator, read the following warnings and read and fully understand
the special conditions of use detailed in the attached certificate of
compliance.
IF UNSURE, CHECK BEFORE USE.
BEFORE USE.
WARNINGS
1.
DO NOT exceed the maximum measurement ratings given on
page 2 of the Certificate of Compliance.
2.
DO NOT open the instrument case in a hazardous area.
3.
Batteries must ONLY be fitted in a SAFE AREA.
4.
If fitted with rechargeable batteries, the batteries must ONLY be
charged in a safe area and only with the charger Type 13603
supplied for use with the Multi-Calibrator.
5.
The RS232 communication circuit may only be used outside the
hazardous area.
6.
When the Multi-Calibrator, Type TRXII IS, is used as a source for
intrinsically safe apparatus, that apparatus may not be connected
to any other intrinsically safe circuit simultaneously. The
parameters of the apparatus must comply with the output
parameters of the Multi-Calibrator.
7.
To prevent electrical shocks or damage to the instrument, do not
connect more than 30 V between the terminals, or between the
terminals and the ground (earth).