K319 Issue 2
EN-9
Englis
h
Changing the temperature scale
Leave the cursor at NONE at the measure and source menu and
press the # and EXE keys simultaneously to open the set-up menu.
Select Temp. settings and press EXE. Select °C and press EXE to
change from °C to °F or vice versa. Press CE to quit the set-up
menu.
Pressure Units Setting
Pressure readings can be shown in different pressure units. The
selection can be made via the Set-up menu.
Changing the pressure units
Leave the cursor at NONE at the measure and source menu and
press # and EXE keys simultaneously to open the set-up menu.
Select PRESSURE UNITS and press EXE. Select one of the 10
different pressure units and press EXE to change. Press CE to leave
the set-up menu.
Note:
The reference temperature for in H
2
0 or mH
2
0 automatically
changes according to the defined temperature setting °C or °F.
°C refers to water temperature at 4°C and °F to 68°F.
Access Code Settings
The ACCESS CODE is programmable by the customer. The
following menus are protected by this code;
•
ACCESS CODE.
•
Calibration menu (Electrical).
•
Add pressure sensors.
•
Delete pressure sensors.
•
Calibration pressure sensors.
Access Code
9410
is the standard factory setting.