EN-28
K319 Issue 2
SPECIAL SOURCE/SIMULATION FUNCTIONS
Step Mode
In the step mode the TRX-II can generate preset source levels in four
different ways;
•
Free programmable, Select PROGR
Number of steps (2 to 10) and levels are free programmable.
•
10% Divisions, Select 10%
Free programmable span is automatically divided in 10 steps of 10%
each.
•
20% Divisions, Select 20%
Free programmable span is automatically divided in 5 steps
of 20% each.
•
25% Divisions, Select 25%
Free programmable span is automatically divided in 4 steps of 25%
each.
Stepping through the fixed levels can either be performed manually
or automatically.
Setting-up programmable steps
Select STEP and PROGR from the menu
and press EXE. Select the desired
number of steps and press EXE again.
Set the source level for each step. Scroll
the step numbers with the
and
keys
to change or enter new levels. Press
EXE when ready. The step mode
working screen is shown.
Setting-up 10%, 20% or 25% division steps
Select STEP and 10%, 20% or 25%
from the menu and press EXE. Set the
desired 100% span and press EXE
again. The step mode working screen
is shown.