DPI 740 User Manual
47
K0200 Issue No. 1
Process commands
Command:
PC - Set-up process
Format:
PC = <process definition>
Description:
This command sets and defines the required
process of the instrument. The <process
definition> must be in the format shown in
the following table:
Process definitions
Example: ~ filter
PC=~(IR,0.15,0.01)
Define filter input reading changing the <time
constant> and <band>.
Note:
This filter does not affect the normal instrument
filter.
PC =<process definition>
Set-up Process
PC =~(IR),<value>,<value>
filter
PC =T(IR),<value>
tare
PC =<(IR)
maximum
PC =>(IR)
minimum
PC =Q(IR),<value>,<value>
QFF
PC =A(IR),<value>
altitude
PR ?
Get process reading
PA =k
Enable automatic sending
of process reading
PM
Reset process maximum
and minimum values.