DPI 740 User Manual
49
K0200 Issue No. 1
Command:
PR - Get reading
Format:
PR?
Description:
Requests the reading.
Query:
This is a query only command.
Example:
OR?
Reply: OR = 1000.00
reading in mbar
Command:
PA - Enable automatic sending of
reading
Format:
PA = k
Description:
Enables automatic sending of input reading.
The value `k` corresponds to `send every `k`th
conversion`. If `k` is 0, the automatic send
feature is disabled.
Example:
PA = 10
send value every 10th conversion.
PA = 0 disable the automatic feature.
Query:
A query responds with an integer of the
automatic send interval in conversions.
Example:PA?
Reply: PA = 10
Command:
PM-
Resets process maximum and
minimum values
(also see PC command `<` and`>`)
Format:
PM
Description:
Sets maximum and minimum values equal
to the current reading, in effect resetting both
values.
Query: There is no corresponding query command.