K0200 Issue No. 1
42
DPI 740 User Manual
example program
#sa?;
query instrument address
!SA=00;
reply address is 00
#fa=1;
put instrument in addressed mode
#0099ic=p;
send command to instrument 00 from
instrument 99 input channel is pressure
#0099pc=~(ir,10,1); set-up process channel to filter 'input
reading ir time constant 10, band 1'
#0099iu=0;
units to mbar
#0099pr?;
get reading from process channel
!9900PR1=987.22;
reply of pressure in mbar (filtered
reading)
#0099ir?;
get unfiltered reading
!9900IR=987.22;
reply of pressure in mbar
#0099iu=18
units to inHg
#0099pr?;
get reading from process channel
!9900PR1=29.153;
filtered pressure reading in inHg
#0099fa=0;
take QFF addressed
#iu?;
query units
!iu=18;
units = 18 = inHg (see table 6-1)
This example is reproduced using QBASIC in Table 6-3.