DPI 740 User Manual
7
K0200 Issue No. 1
RS232 connections
Option C is the adaptor cable recommended for use with the
RS232 serial communication interface. The adaptor cable has,
at one end, 6-pin LEMO type connector and, at the other, a 9-
way D type connector. The instrument's serial communication
settings and the data terminal equipment (DTE) settings must
be the same. The recommended initial settings are as follows:
Baud rate
-
9600
Data bits
-
8
Stop bits
-
1
Parity
-
none
Handshaking
-
none
Adaptor Cable
Notes
1. The function column for the D type connector uses RS232
terminology with respect to DTE.
2. In the D-type connector DTR and DSR should be connected
together as follows:
9-way
pin 4 and 6
25-way
pin 20 and 6
3. The instrument's serial communication settings can be
changed using the set-up menu.
Pin No.
Function
Flow
Function
D type Connector
1
2
3
5
6
4
RxD input
CTS input
GND (screen)
not used
TxD output
RTS output
TxD
RTS
GND
-
RxD
CTS
3
7
5
-
2
8
25-way
9-way
Pin No.
Pin No.
2
4
7
-
3
5
LEMO connector
Instrument (DCE)
PC (DTE)
Summary of Contents for Druck DPI 740
Page 1: ...GE Sensing Druck DPI 740 Precision Pressure Indicator User Manual K0200 ...
Page 2: ... The General Electric Company All rights reserved ...
Page 16: ...K0200 Issue No 1 8 DPI 740 UserManual Figure 2 2 Electrical connections ...
Page 32: ...K0200 Issue No 1 24 DPI 740 User Manual intentionally left blank ...
Page 46: ...K0200 Issue No 1 38 DPI740 UserManual ...
Page 72: ...K0200 Issue No 1 64 DPI 740 UserManual Approved Service Agents www gesensing com ...