5 Operation
Carrying out the calibration
5-28
Edition 4 (05/2014)
DMS Go+
1-point calibration
The 1-point calibration is carried out using the reference
block at the instrument rear and a calibration block with
known thickness which is made of the same material as
the test object.
Ideally, the wall thickness of the calibration block is the
equal to or slightly greater than the expected readings
from the test object.
– In the function group
PROBE&CAL
, select the func-
tion
THK CAL
.
– Choose the option
1-PT
.
– In the function group
PROBE&CAL
, select the func-
tion
ZERO
.
– Select the method
MANUAL
.
– In the function group
PROBE&CAL
, select the func-
tion
CALIBRATION
.
– Press one of the
left
or
right
arrow keys in order to
start the calibration. The message
Put probe on
zero block - use couplant
is displayed on the
screen.
– Apply some couplant to the instrument reference
block and couple the probe. The zeroing is started.
After finishing it, the message
Probe zero complete.
Remove probe from zero block.
is displayed on the
screen.
– Uncouple the probe. The message
Couple to Cal
standard (displayed below) or press ENTER to
change cal. block thickness.
is displayed on the
screen.
– If the wall thickness displayed on the screen is not
identical with the thickness of the calibration block,
press the function key 2 (function
ENTER
).
– Select the digits that you wish to change using the
left
and
right
arrow keys, and change the value using
the
up
and
down
arrow keys.
– Press the function key 2 (function
DONE
) in order to
close the process of changing the thickness value.
– Apply some couplant to the calibration block with
known thickness and couple the probe.
Summary of Contents for DMS Go+
Page 1: ...DMS Go Technical Reference and Operating Manual ...
Page 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Page 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Page 8: ...0 8 Edition 4 05 2014 DMS Go ...
Page 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Page 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Page 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Page 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Page 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Page 78: ...4 12 Edition 4 05 2014 DMS Go ...
Page 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Page 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Page 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Page 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Page 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Page 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Page 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Page 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Page 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Page 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Page 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Page 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Page 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Page 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Page 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Page 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Page 252: ...11 10 Edition 4 05 2014 DMS Go ...