Preparing for measurements
5 Operation
DMS Go+
Edition 4 (05/2014)
5-23
DUAL
The measurement mode DUAL is available for dual-el-
ement probes.
The wall thickness is determined on the basis of the ul-
trasonic time of flight from zero up to the last zero cross-
ing before the first flank crossing detected in gate A. The
ultrasonic time of flight is corrected by the probe delay
line which is calculated automatically during the zero
point calibration.
This measurement mode is suitable for non-coated ma-
terials showing corrosion at the rear, especially with
curved and irregular backwall or inner wall surfaces.
DUAL-MULTI
The measurement mode DUAL-MULTI is available for
dual-element probes.
The wall thickness is determined on the basis of the ul-
trasonic time of flight from the last zero crossing before
a peak echo is detected in gate A and ends with the last
zero crossing before a peak echo in gate B.
This measurement mode is suitable for coated materials
with smooth, parallel surfaces.
Note
Only the measurement modes supported by
the selected probe type are displayed as op-
tions to choose from. The measurement
modes
TOP-COAT
and
AUTO-V
are only
available for DMS Go+ TC and described in
detail in chapter 7 of this manual.
ATTENTION
The choice of measurement mode must al-
ways be identical for the calibration and the
subsequent measurements.
– In the function group
PROBE&CAL
, select the func-
tion
TG MODE
.
– Select the suitable measurement mode for the con-
nected probe and measurement task. The instrument
is immediately adjusted.
Summary of Contents for DMS Go+
Page 1: ...DMS Go Technical Reference and Operating Manual ...
Page 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Page 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Page 8: ...0 8 Edition 4 05 2014 DMS Go ...
Page 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Page 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Page 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Page 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Page 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Page 78: ...4 12 Edition 4 05 2014 DMS Go ...
Page 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Page 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Page 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Page 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Page 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Page 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Page 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Page 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Page 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Page 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Page 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Page 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Page 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Page 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Page 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Page 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Page 252: ...11 10 Edition 4 05 2014 DMS Go ...