5 Operation
Preparing for measurements
5-22
Edition 4 (05/2014)
DMS Go+
Choosing the measurement mode
You have to choose the suitable measurement mode for
the task. By doing so, you define the method for operat-
ing the instrument or the data used for determining the
measurement result.
You have a choice between the following measurement
modes:
S-IP
The measurement mode S-IP is available for single-el-
ement probes.
The wall thickness is determined on the basis of the ul-
trasonic time of flight from zero up to the last zero cross-
ing before the first flank crossing detected in gate A. The
ultrasonic time of flight is corrected by the probe delay
line which is calculated during the probe zero offset (see
Section
This measurement mode is suitable for non-coated ma-
terials with smooth, parallel surfaces.
S-PEAK
The measurement mode S-PEAK is available for single-
element probes.
The wall thickness is determined on the basis of the ul-
trasonic time of flight from the last zero crossing before
the amplitude peak in gate A up to the last zero crossing
before the peak echo in gate B.
This measurement mode is suitable for non-coated ma-
terials with smooth, parallel surfaces.
S-FLANK
The measurement mode S-FLANK is available for sin-
gle-element probes.
The wall thickness is determined on the basis of the ul-
trasonic time of flight from the last zero crossing before
the first flank detected in gate A up to the last zero
crossing before the first flank crossing detected in
gate B.
This measurement mode is suitable for coated materials
with smooth, parallel surfaces.
Summary of Contents for DMS Go+
Page 1: ...DMS Go Technical Reference and Operating Manual ...
Page 3: ...DMS Go Edition 4 05 2014 0 3 Function groups and functions ...
Page 4: ...0 4 Edition 4 05 2014 DMS Go Function groups and functions cont d ...
Page 8: ...0 8 Edition 4 05 2014 DMS Go ...
Page 17: ...DMS Go Edition 4 05 2014 1 1 Introduction 1 ...
Page 41: ...DMS Go Edition 4 05 2014 2 1 Standard package and accessories 2 ...
Page 54: ...2 Standard package and accessories Recommended accessories 2 14 Edition 4 05 2014 DMS Go ...
Page 55: ...DMS Go Edition 4 05 2014 3 1 Initial start up 3 ...
Page 67: ...DMS Go Edition 4 05 2014 4 1 Principles of operation 4 ...
Page 78: ...4 12 Edition 4 05 2014 DMS Go ...
Page 79: ...DMS Go Edition 4 05 2014 5 1 Operation 5 ...
Page 133: ...DMS Go Edition 4 05 2014 6 1 Data Recorder 6 ...
Page 174: ...6 Data Recorder Documenting measurements 6 42 Edition 4 05 2014 DMS Go ...
Page 175: ...DMS Go Edition 4 05 2014 7 1 Special functions of the DMS Go TC 7 ...
Page 189: ...DMS Go Edition 4 05 2014 8 1 Maintenance and care 8 ...
Page 195: ...DMS Go Edition 4 05 2014 9 1 Interfaces and Peripherals 9 ...
Page 199: ...DMS Go Edition 4 05 2014 10 1 Appendix 10 ...
Page 228: ...10 Appendix Recycling directives 10 30 Edition 4 05 2014 DMS Go 4 5 1 4 ...
Page 231: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 33 2 ...
Page 234: ...10 Appendix Recycling directives 10 36 Edition 4 05 2014 DMS Go 1 3 ...
Page 235: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 37 1 1 ...
Page 236: ...10 Appendix Recycling directives 10 38 Edition 4 05 2014 DMS Go 5 ...
Page 237: ...Recycling directives 10 Appendix DMS Go Edition 4 05 2014 10 39 4 7 ...
Page 238: ...10 Appendix Recycling directives 10 40 Edition 4 05 2014 DMS Go 3 6 6 6 7 ...
Page 242: ...10 Appendix Recycling directives 10 44 Edition 4 05 2014 DMS Go ...
Page 243: ...DMS Go Edition 4 05 2014 11 1 Specifications 11 ...
Page 252: ...11 10 Edition 4 05 2014 DMS Go ...