EMC Compliance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
C-3
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Note:
U
T
is the AC mains voltage prior to application of the
test level.
Conducted RF
IEC 61000-4-6
3
VRMS
150 kHz to 80 MHz
3
VRMS
150 kHz to 80 MHz
Portable and mobile RF communications equipment should be
used no closer to any part of the system, including cables,
than the recommended separation distance calculated from
the equation appropriate for the frequency of the transmitter.
For recommended separation distances, see
Radiated RF
IEC 61000-4-3
(alternative method:
IEC 61000-4-21)
3
V/m
80 kHz to 2.5 GHz
3
V/m
80 kHz to 2.5 GHz
*
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects,
and people.
†
UT is the AC mains voltage prior to application of the test level.
Table C-2: EMC Immunity Guidance and Declaration (Continued)
*
Immunity Test
IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment Guidance