Equipment Description and General Construction Requirements
2.2 Room Size, Layout and Considerations
2.2.3 System Mechanical Curves
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
2-21
2.2.3 System Mechanical Curves
Figure 2-14: D670 Component Movement Curves, p.2-22
illustrates the table and gantry
movement. In addition, the contrast injector, ECG trigger monitor and collimator carts can be
moved to different locations in the scan room, as demonstrated in the layout illustrations in
through
NOTE
In order to prevent collision with the gantry display boom, do not mount any equipment from the ceiling.
Table slanted at
Farthest point relative to system's center line
67.5
°
190.5 cm (75")
55
°
181.5 cm (71.5")
42.5
°
157.5 cm (62")