General System Requirements
1.3 Delivery Requirements
1.3.4 Crated and Uncrated Weights, Measurements and Clearance
Discovery NM/CT 670 – Standard Integration Instructions
5483063-1EN, Rev.2, ©2015 GE Healthcare
1-13
Table 1-3: D670 Components and Clearance — Imperial
Crated
Uncrated
Component name
Crate size
(
"
)
(Dollies not attached)
(Height × Width ×
Depth/Length)
Weight
(lb)
Minimal dimensions
(")
*
Weight
(lb)
Door
width
Corridor /
elevator
width
Corridor /
elevator
length
Width of
corridors with
90
°
turns
†
Height
Pre-installation kit
‡
29.5×5.7×68.9
33
any
any
any
any
any
33
NM gantry
86.6×59×66.1
4917
55.1
55.1
88.6
98.4
78.75
4531
(with dollies)
Table
55×35.4×118.1
1625
39.4
39.4
111.4
98.4
any
1245.6
NM acquisition station
31.5×23.62×23.62
66
any
any
any
any
any
44
Peripherals and accessories
45.3×39.4×59
110
any
any
any
any
any
110
Collimators on carts
67×35.4×45.3
816
(heaviest
coll. set)
22
22
39.4
45.3
59
727.5
(heaviest coll.
set)
CT gantry
91x54x78.75
4740
45.3
45.3
110.6
98.4
78.75
4432
(with dollies)
NM/CT covers
81.5×41.3×77.6
198
any
any
any
any
any
55
Optional Items
NM UPS
May vary but not
more than
23.6x15.7x31.5
May vary
but not
more
than
330.7
any
any
any
any
any
May vary
but not
more than
286.6