Figure 2-20:D670 Floor Anchor Points
Table 2-3: D670 Drilling and Anchor Chart
No.
X
Y
Drill
Hole
Hole
Depth
Anchored
Part
Hole Purpose
Drilling
Method
Anchor
Type
Torque
Nm.
Section
1
-0.00
405.00
Ø15.0
113.0
NM gantry
Main Anchor
Metal
Drilling
Template
HILTI
HSL-3-G
M10/100
35
2
-0.00
-405.00
3
-810.00
405.00
4
-810.00
-405.00
5
-80.00
365.00
Alternative Anchor
6
-80.00
-365.00
7
-810.00
530.00
8
-810.00
-530.00
NM gantry anchor points
4 × HILTI-HSL-3-G M10/100
anchors
CT gantry anchor points
8 × HILTI-HSL-3 M10/20
anchors
Table anchor plate