background image

PRO Series 100 - 400V2 
Pre-Service and 
Calibration Manual

(Models 110 - 410)

Summary of Contents for Dinamap PRO Series

Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410...

Page 2: ......

Page 3: ...The Model Numbers 100 through 400 are generic in nature and reflect the range of product codes available Your product may be labeled with a specific product code such as DINAMAP PRO Model 410 V2 refe...

Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...

Page 5: ...58 5120 US only Fax 414 355 3790 Part No 2018554 001 A The content of this document including all figures and drawings is proprietary information of GE Medical Systems Information Technologies provide...

Page 6: ...800 558 5120 US only Fax 414 355 3790 European Representative GE Medical Systems Information Technologies GmbH Munzinger Stra e 3 5 D 79111 Freiburg Germany Tel 49 761 45 43 0 Fax 49 761 45 43 233 Asi...

Page 7: ...ration 4 1 4 Clinician Mode 4 1 4 1 Accessing the Clinician Mode 4 1 4 2 Clinician Mode Main Menu 5 1 4 3 Clinician More Menu 6 1 5 Service Mode 8 1 5 1 Accessing the Service Mode 8 1 5 2 Service Mode...

Page 8: ...Revision A 2018554 001 Table of Contents 1 7 1 Calibrating the NIBP Transducer 20 1 7 2 Overpressure Adjustment 21 1 8 Electrical System Safety Checks 23 1 9 Features Selections 23 1 10 Troubleshootin...

Page 9: ...ower to mains power when it is plugged in to an AC power source To confirm a fuse problem remove the battery plug the unit in and turn it on If the PRO Monitor does not operate the probable cause is a...

Page 10: ...up Confirm that all of the pixels on the LCD display are lit momentarily and the backlight is powered during initialization The backlight is only noticeable in a dimly lit room Confirm that the displ...

Page 11: ...er indicator is lit and the battery LED is off 2 Remove the battery and confirm that external DC LED on the Monitor flashes 3 Fit the battery and disconnect the external power supply Confirm the exter...

Page 12: ...a special mode named Clinician Mode This mode is accessed from the More option on the Main menu and requires a code before a user can access it The Clinician Mode enables the user to Verify NIBP cali...

Page 13: ...ual buttons on this menu are described briefly below Press Button Selecting the Press button displays the target pressure dialog box pictured above which sets the default target inflation pressure for...

Page 14: ...ally clear trend data on power up The default setting is Yes In the default setting trend data is cleared when the PRO Monitor is shut off Select No to retain the trend data on power down Print Button...

Page 15: ...ings The illustration above shows the default settings The user can adjust the alarm limits within the ranges specified in the table below The high limit must be at least one step higher than the low...

Page 16: ...Re calibrate BP Adjust the pneumatic system overpressure point Configure communications with a host computer Change the language of operation 1 5 1 Accessing the Service Mode To enter Service Mode fro...

Page 17: ...most recent NIBP calibration and PM check Select OK to return the display to the Service menu System Button This button accesses the System menu The options on this menu set the display language commu...

Page 18: ...and the calibration procedure Cal Button This button initiates the calibration procedure See Section 4 7 for a step by step description of this procedure Pop Off Button This button functions in a simi...

Page 19: ...will be displayed after the power up screen informing the user that preventative maintenance is due This screen can be dismissed by pressing any front panel key and normal operation will begin The dat...

Page 20: ...saves the selection and a dialog box requests that the Monitor be turned off If no language is selected when OK is pressed the language is stored as undefined In this case the Monitor prompts the user...

Page 21: ...ns are also listed on the error log ID Tag Button Selection of the ID Tag button opens a menu that allows the user to enter a 12 digit Monitor Asset Tag for their own use If the Asset ID Tag is not eq...

Page 22: ...the sensor interfaces or by measuring of specific parameters The test procedures employ features of the operational software and test modes of the service menus A guide to the Monitor controls indica...

Page 23: ...ellcor SpO2 system For Monitors equipped with Masimo SpO2 Masimo recommends BIO TEK SpO2 simulators 1 Connect the appropriate SpO2 simulator and cable to the SpO2 connector 2 Verify the unit displays...

Page 24: ...ter 2 Ensure the index line of the CRITIKON Blood Pressure Cuff is properly aligned within the range markers on the opposite end of the cuff 3 Power on the PRO Monitor and select More 4 Select the Ser...

Page 25: ...ecessary to have the following test equipment 2 tube NIBP hose 12ft p n 107365 DINAMAP Adult BP cuff p n 2774 or equivalent Calibration Kit p n 320246 with a manual inflation bulb NIST calibrated sing...

Page 26: ...the Service mode and input 2 2 1 3 4 From the Service menu select the NIBP button then select the Popoff button Note If the overpressure point is out of range adjust the overpressure potentiometer as...

Page 27: ...ogies does not stock this tester To check the temperature system connect the IVAC probe simulator to the temperature probe connector on the front panel and insert a temperature probe into the active h...

Page 28: ...tor as shown in Figure 4 1 The following procedure describes the steps required to calibrate the pressure transducer Disassembly is not required 1 7 1 Calibrating the NIBP Transducer 1 Set up the Moni...

Page 29: ...correct pressures are applied at these two points After venting the Monitor displays Calibration is Complete or Calibration Failed If the calibration fails turn off the Monitor check the calibration...

Page 30: ...he pressurized system 6 Select the Check button again and verify that the overpressure trips at 315 mmHg 10 mmHg 7 Select the Pop off button This button sets the overpressure point to the neonatal val...

Page 31: ...s voltage applied to the test probes Use an SpO2 plug with the leads shorted together and apply the test probe of the analyzer to the leads The leakage reading should be less than 150 A 1 9 FEATURES S...

Page 32: ...A 2018554 001 contains both a 9 and an x Select OK and a tone sounds signalling that your change has been made c Clear Error Log Selecting this button clears all errors in the error log and sounds a...

Page 33: ...has ceased operation Negative supply to LCD 9 6 Vdc is out of tolerance Replace LCD Replace main board Check negative supply reaching the LCD module No LCD backlight LCD faulty Main board fault Replac...

Page 34: ...bored generally accompanied by one of the following messages N33 NIBP Inflation Timeout N00 NIBP Overpressure Reversed connections on pneumatic assembly failure Blocked or kinked hose from pump output...

Page 35: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual 27 2018554 001 1 11 ERROR CODES...

Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001...

Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix...

Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001...

Page 39: ...reading at 100 saturation 96 100 5 3 1 5 SP02 reading at 90 saturation 86 94 5 3 1 6 SP02 reading at 80 saturation 76 84 5 5 1 7 Verify alarm sounds and displays below 90 5 3 2 BPM test 5 3 2 1 Verify...

Page 40: ...erify battery indicator 5 4 5 Remote Alarm 5 4 5 2 Voltage between pins 4 and 8 alarm inactive 4 7 5 3 5 4 5 4 Voltage between pins 4 and 8 alarm active 0 0 0 1 5 4 6 Communications test 5 4 6 2 Verif...

Page 41: ......

Page 42: ...GE Medical Systems Information Technologies GmbH Munzinger Stra e 3 5 D 79111 Freiburg Germany Tel 49 761 45 43 0 Fax 49 761 45 43 233 Asia Headquarters GE Medical Systems Information Technologies Asi...