1.6.7 Printer Test
This test generates a sample printout from the printer. If no paper is in
the printer, the Monitor generates no alarm.
1. Power on the PRO Monitor and select
More...
2. Select the
Service
button and input
8 3 7 8
.
3. From the Super Service Menu, select the
button,
Ensure the printed test page is clear and easy to read.
1.7 PRO MONITOR CALIBRATION PROCEDURES
Calibration procedures include calibration of the transducers and
adjustment of neonatal and adult overpressure points. These tests
require a manometer and an inflation bulb attached to the PRO
Monitor as shown in Figure 4-1.
The following procedure describes the steps required to calibrate the
pressure transducer. Disassembly is not required.
1.7.1 Calibrating the NIBP Transducer
1. Set up the Monitor and calibration equipment as shown in Fig. 4-1.
2. Power on the PRO Monitor.
3. Verify that calibration equipment reads 0 mmHg of pressure.
Note:
If the PRO Monitor displays the language choice menu, select
a language. Select
OK
and reboot the system.
4. Select the
More…
button from the Main menu.
5. Select the
Service
button.
6. Enter the Service Mode access code
2 2 1 3
.
7. Select the
NIBP
button from the Service menu.
Important:
From this point, the timing is critical. Before proceeding,
review the following steps.
Setting the Calibration Points
8. Select
Cal
from the NIBP menu.
9. The Monitor displays “Set Pressure to 0 mmHg.”
10. Ensure calibration test equipment is at 0 mmHg.
Summary of Contents for Dinamap PRO 100
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Page 41: ......