A-4
DINAMAP® PRO Series 100-400V2 Monitor Service Manual
Revision A
2018554-001
Appendix A
Step
Description
Min
Max
Actual Pass Fail N/A
5.3.4.5
Measured temperature in
°
F (80.2 nominal)
79.9 80.5
5.3.4.7
Measured temperature in
°
F (107.8 nominal)
107.5 108.1
5.3.4.10 Broken Probe fault condition
5.4
Pushbutton test
5.4.1.1
Verify appropriate responses to pushbuttons
5.4.2
Sound test
5.4.2.2
Verify volume levels
5.4.3
Printer test
5.4.3.2
Printer produces a clear printout
5.4.4
Battery System test
5.4.4.1
Verify AC mains indicator
5.4.4.4
Verify uninterrupted battery operation
5.4.4.5
Verify battery indicator
5.4.5.
Remote Alarm
5.4.5.2
Voltage between pins 4 and 8, alarm inactive
4.7 5.3
5.4.5.4
Voltage between pins 4 and 8, alarm active
0.0 0.1
5.4.6
Communications test
5.4.6.2
Verify pump starts, stops and temp. status returns
6.0
Hipot Tests
6.1
AC Main Hipot
6.2 Temp
Hipot
6.3
SP02 Hipot
7.0
Safety Tests
7.2.7
EXTDC to GND Resistance (m
Ω
)
0 1000
7.3.6
SPO2 leakage (uA) (models 300 and 400)
0
150
7.4.6
Temp leakage (uA) (models 200 and 400)
0
50
7.5.4.1
Normal no-fault leakage (uA)
0
500
7.5.4.3
Normal open-ground leakage (uA)
0
500
7.5.4.6
Normal open-neutral leakage (uA)
0
500
7.5.5.1
Reverse no-fault leakage (uA)
0
500
7.5.5.3
Reverse open-ground leakage (uA)
0
500
7.5.5.6
Reverse open-neutral leakage (uA)
0
500
Tested By: ______________________________________ Date: ___________
Summary of Contents for Dinamap PRO 100
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Page 41: ......