16
DINAMAP® PRO Series 100-400V2 Monitor Pre-Service & Calibration Manual
Revision A
2018554-001
1.6.2 Leak Test
This test performs a leak test of the pneumatic system.
Equipment required:
CRITIKON
®
Adult Blood Pressure Cuff (P/N 2774)
12 foot Gray, Adult/Pediatric Air Hose (P/N 107365)
Figure 4-1
1. Secure the Adult-size cuff to the 12-foot air hose. Attach to the
PRO Monitor as shown in Figure 4-1. Secure the BP cuff around a
rigid, unbreakable object that measures at least 2 inches in diameter.
2. Ensure the index-line of the CRITIKON
®
Blood Pressure Cuff is
properly aligned within the range-markers on the opposite end of the cuff.
3. Power on the PRO Monitor and select
More...
4. Select the
Service
option and input
8 3 7 8.
5. From the Super Service menu, select the
NIBP
option.
6. From the NIBP menu, select the
LEAK
option.
The leak-test sequence closes both valves and turns on the pump.
The PRO Monitor will self-pressurize the pneumatic test setup to
approximately 200 mmHg. After settling for 45 seconds, the test
pressure value will be displayed in mmHg on the LCD. After 30
seconds, the system displays the final pressure. Confirm that the
pressure has fallen no more than 5 mmHg.
Summary of Contents for Dinamap PRO 100
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Page 41: ......