1.8 ELECTRICAL SYSTEM SAFETY CHECKS
The PRO Monitor is designed to protect the patient from electric
shock. To ensure the integrity of these safeguards, use a safety
analyzer to perform the following leakage current tests. These tests
should be performed on every unit whenever the case is opened
before it is returned to clinical service. If any reading exceeds the limit
specified, do not return the unit to clinical service.
Ground Leakage
Measure ground leakage current with normal polarity, reverse
polarity, and open neutral. Limit is 500 µA.
Temperature System Leakage (PRO 200 & 400)
Set your analyzer to measure leakage current with mains voltage
applied to the test probes. Use a temperature plug with the leads
shorted and apply the test probe of the analyzer to the leads. The
leakage reading should be less than 150 µA.
SpO
2
System Leakage (PRO 300 & 400)
Set your analyzer to measure leakage current with mains
voltage applied to the test probes. Use an SpO
2
plug with the
leads shorted together and apply the test probe of the analyzer
to the leads. The leakage reading should be less than 150 µA.
1.9 FEATURES SELECTION
The PRO Monitor allows the user to turn features (e.g., SpO
2
, Temp,
Printer) on and off. To access this option:
1. Power on the PRO Monitor and select the
More...
button.
2. Select
Service
mode and enter
3122
.
3. Select the feature button you wish to change:
a.
Select SpO
2
: (
NELLCOR, MASIMO,
or
NONE
) Selecting
this button allows you to turn the SpO
2
parameter on or off. Choose
the type of SpO
2
technology (Nellcor or Masimo) that is in your
Monitor to turn it on or off by using the rotor to select the option.
Once you have made your selection, a small check mark
9
appears next to the box that contains both a
9
and an” x.” If your
Monitor does not have an SpO
2
parameter, the
NONE
box should
be selected. Select
OK
and a tone sounds signalling that your
change has been made.
b.
Feature Set Up
: (
Temp
and
Printer
) Selecting this button
allows you to turn the Temp parameter and printer on or off by
using the rotor to select the option. Once you have made your
selection, a small check mark
9
appears next to the box that
Summary of Contents for Dinamap PRO 100
Page 1: ...PRO Series 100 400V2 Pre Service and Calibration Manual Models 110 410 ...
Page 2: ......
Page 4: ...DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 36: ...28 DINAMAP PRO Series 100 400V2 Monitor Pre Service Calibration Manual Revision A 2018554 001 ...
Page 37: ...Revision A DINAMAP PRO Series 100 400V2 Monitor Service Manual A 1 2018554 001 A Appendix ...
Page 38: ...A 2 DINAMAP PRO Series 100 400V2 Monitor Service Manual Revision A 2018554 001 ...
Page 41: ......