Appendix C. Factory Tests
64
DigitalFlow™ XGM868i Service Manual
C.3 Resetting the XGM868i to Factory Defaults (cont.)
7.
Scroll to
LCD
and press
[Enter].
8.
The window now asks for the
# of LCD Parameters
. Scroll to the desired number (from OFF through 1-4 and
KEY) and press
[Enter].
For a 1-Channel XGM868i, the data for Channel 1 is displayed automatically, and you can skip to step 9. However, for
a 2-Channel meter, the channel data to be displayed must be specified at the following prompt.
9.
Scroll to the desired
Channel option
10.
For each channel, select the desired
Measurement Parameter
, as shown in the
Programming Manual.
Note:
The measurement units that appear in these prompts are those selected in the
GLOBL-SYSTM
menu.
The previous two prompts repeat until all of the specified
# of LCD Parameters
have been set up. When all of the
display parameters have been set up, the meter returns to the
Global I/O
window. To leave the
Keypad Program
, press
the
[Escape]
key three times.
After leaving the
Keypad Program
, the XGM868i will reset itself and will begin to display the parameters specified in
this section. If more than one parameter was set up, each of the parameters will be displayed in sequence, with a pause
of several seconds between display changes.
Table 11: Channel Options
Option
Description
CH1
Channel 1
CH2
Channel 2
SUM
CH1+CH2
DIF
CH1-CH2
AVE
(CH1+CH2)/2
Summary of Contents for DigitalFlo XGM868i
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 44: ...Chapter 4 Parts Replacement DigitalFlow XGM868i Service Manual 36 ...
Page 48: ...40 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 70: ...62 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 78: ...Warranty 70 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 79: ......