Appendix B. Calibration and Testing with PanaView
B.4.2 Analog Inputs
Analog inputs may be added to the Model XGM868i by installing an appropriate option card in
Slot 1
. This option card
contains two or four analog inputs, which are designated as A, B, C and D. Both the zero-point and full-scale values for
each input must be calibrated.
Calibration of the analog inputs requires the use of a calibrated current source. If an independent calibrated current
source is not available, one of the
Slot 0
analog outputs may be used for the calibration. During the analog input
calibration, the
Slot 0
analog output will supply the low reference, high reference, 4 mA and 20 mA signals at the
appropriate times.
IMPORTANT:
If a
Slot 0
analog output will be used to calibrate the analog inputs, the
Slot 0
analog output calibration
procedure must be completed first.
Prepare for the calibration procedure by connecting a
Slot 0
analog output (or independent calibrated current source) to
Input A
of the option card. See Figure 4 on page 12 to identify the
IN(+)
and
RTN(-)
pins for
Input A
on terminal block J2.
Note:
The zero point of the analog input may be set for either 0 mA or 4 mA. However, the calibration procedure
always uses the 4 mA point, as the meter will extrapolate this value to obtain the 0 mA point.
1.
In the center pane of the
Calibration/Test
window, double-click on
Slot 1
.
2.
Double-click on the desired input to open the
Analog Input
menu. (
Input A
will be used as an example in this
manual.)
IMPORTANT:
Calibrating
Inputs B, C or D
is identical to calibrating
Input A
. However, be sure to reconnect the current
source on terminal block
J2
. Refer to Figure 4 on page 12 for the correct pin numbers.
3.
Double-click on the appropriate option to select the reference point to be calibrated.
Proceed to the corresponding sub-section for specific instructions on each of the options available in step 3.
B.4.2a 4 mA Option
Set the calibrated current source to 4 mA.
•
If you selected the
4 mA
option in the
Analog Input
menu, double-click on
Store
to accept the current 4 mA value or
on
Abort
to cancel the entry. In either case, the
Analog Input
prompt reappears.
B.4.2b 20 mA Option
Set the calibrated current source to 20 mA.
•
If you selected the
20 mA
option in the
Analog Input
menu, double-click on
Store
to accept the current 20 mA value
or on
Abort
to cancel the entry. In either case, the
Analog Input
prompt reappears.
Click
[Exit Page]
to return to the
Slot 1
window and calibrate additional inputs/outputs, or click
[Exit Page]
again to
close the
Calibration/Test
menu.
Summary of Contents for DigitalFlo XGM868i
Page 2: ......
Page 4: ...ii no content intended for this page ...
Page 44: ...Chapter 4 Parts Replacement DigitalFlow XGM868i Service Manual 36 ...
Page 48: ...40 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 70: ...62 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 78: ...Warranty 70 DigitalFlow XGM868i Service Manual no content intended for this page ...
Page 79: ......