Parts Replacement
Installing an Option Card
The CTF878 flowmeter can accommodate up to six option cards. The
option cards are installed into sockets on the rear of the printed circuit
board, and they are held in place with a metal bracket. A single metal
bracket is used to secure all the installed option cards.
Note:
If the CTF878 presently has no option cards installed, be sure
to purchase the metal mounting bracket along with the first
option card.
To install an option card, refer to Figure 5-1 on page 5-9 and
Figure 5-2 on page 5-10, and complete the following steps:
1.
Remove the printed circuit board, as described on page 5-3.
2.
Place the printed circuit board face down on a clean, flat surface.
3.
If one or more option cards are already installed, remove the four
fasteners that secure the metal bracket to the printed circuit board.
Lift the metal bracket straight up and away from the printed circuit
board.
Note:
The fasteners may be discarded, as new plastic snap rivets
will be provided.
4.
There are six 32-pin option card sockets (
J41–J46
) on the rear of
the printed circuit board. To install an option card, insert its 32-pin
connector into any available option card socket and gently press
the card into place. Make sure that the pins in the connector are
straight and properly aligned with the socket and that the
connector is positioned on the right side of the option card.
Caution!
Do not force the option card into the socket. If the card
does not enter the socket easily, check for and straighten
any bent pins in the connector and try again.
5.
Repeat step 4 to install any additional option cards.
6.
Place the metal bracket over the option cards, making sure that all
option cards are aligned with the plastic card guides in the bracket.
Secure the metal bracket to the printed circuit board with the snap
rivets provided. See the installed assembly in Figure 5-1 on
page 5-9 and Figure 5-2 on page 5-10.
Complete the option card installation by proceeding to
Installing the
Printed Circuit Board
on page 5-5.
Summary of Contents for CTF878
Page 1: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter User s Service Manual ...
Page 2: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter User s Service Manual 910 254SA4 February 2009 ...
Page 6: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 43: ...Chapter 4 ...
Page 51: ...Chapter 5 ...
Page 63: ...Appendix A ...
Page 64: ...Service Record Introduction A 1 Data Entry A 1 ...