Calibration
3-9
February 2009
Calibrating RTD’s
The RTD calibration procedure consists of setting the low and high
temperature readings.
To calibrate an RTD you must connect an RTD to the RTD board.
Proceeding as above, select the desired Calibrate RTD option, and
press [
ENTER
]. The screen appears as in Figure 3-8 below. Step
through options using the arrow keys.
Figure 3-8: RTD Calibration Window
1.
The first prompt asks you to set the low temperature.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your entry.
2.
Press the [
X
] arrow key to move to the Read box, and then press
[ENTER]
to read the low temperature.
3.
Scroll to the Calibrate box and press [
ENTER
]. The low point has
been calibrated.
4.
The next prompt asks for the value to which you want to set the
high temperature.
a.
Press
[ENTER]
to open the text box.
b.
Use the numeric keys to enter the desired value.
c.
Press
[ENTER]
to confirm your entry.
5.
Press the [
X
] arrow key to move to the Read box, and then press
[ENTER]
to read the high temperature.
6.
Scroll to the Calibrate box and press [
ENTER
]. The high point has
been calibrated.
Calibrate Analog Input
4
Totalizer/Frequency 3:0
1000
OK
Cancel
C
Calibrate
Reset
Low
High
100 C
Read
Read
Calibrate RTD
0
Summary of Contents for CTF878
Page 1: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter User s Service Manual ...
Page 2: ...GE Sensing Model CTF878 Clamp On Tag Flowmeter User s Service Manual 910 254SA4 February 2009 ...
Page 6: ...Chapter 1 ...
Page 18: ...Chapter 2 ...
Page 30: ...Chapter 3 ...
Page 43: ...Chapter 4 ...
Page 51: ...Chapter 5 ...
Page 63: ...Appendix A ...
Page 64: ...Service Record Introduction A 1 Data Entry A 1 ...