System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
42
1-4-3
Mis Chart
System MPH 50/65/80 and RS85 Tube Stand with Compax 40E / SG 60 or SG100
Note:
Some of the cables, or set cables, are factory connected at one end. In the following MIS Charts:
−
Numbers on the left presented in grey squares are HLA numbers (CABLES #).
−
The
symbol is used to identify those connections which remain to be done during on site
installation.
−
The
symbol indicates where cable clamping is necessary (refer to Job Card IST006 of this Manual
for EMC rules requirements.
DO NOT FORGET TO CLAMP CABLES ENTERING THE MPH CABINET. IT IS
MANDATORY TO MAINTAIN THE SYSTEM EMC COMPLIANT.
HV TANK
X−RAY TUBE
1
2145653
ST1−1
ST1−3
ST1−5
ST1−10
ST1−11
Clamped
ST1−7
Black
White
Green
Brown
Blue
Orange
2
3
X1−L
X1−N
PE1−GND
2165548
4
STATOR CABLE
RS85 Power Supply CABLE
3600208..0/1/2/8
HV
TERMINALS
X−RAY TUBE
(Maxiray−75/100)
MPH A1 A2 MODULE
(Stator Connector Panel)
PRINTER−X Power Supply CABLE
ST1−12
Yellow
Shield
2104204
XJ1A/B−1
TSxx−L
TSxx−N
GND STUD
MPH 50/65/80 CABINET
PRINTER − X
MPH A9 MODULE
MPH 50/65/80 CABINET
Black
White
Green
Brown
Blue
Orange
Yellow
Shield
GND Screw
(1)− Pressure Switch
(2)−Pressure Switch
(3)−Terminal Strip
(5)−Terminal Strip
(6)−Terminal Strip
(7)−Terminal Strip
(8)−Terminal Strip
MPH A6 A3 MODULE
(Distribution Board)
XJ1A/B−2
Ground Plug
MPH A1 MODULE
(AC Distribution Panel)
F15−2
F14−2
Ground Plug
120/240Vac
0Vac
Ground
120/240Vac
0Vac
Ground
0Vac
230Vac
Gnd
MIS 107
MIS
27239
MIS 27238
MPH SCI
RS 85 POWER
SUPPLY
W34
CABLE #
TUBE−I (+)
ANODE
ANODE
TUBE−I (−)
CATHODE
CATHODE
X−RAY TUBE HV
CONNECTIONS
L
L
C
S
M
C
S
M
L
L
C
S
M
C
S
M
3600208..0/1/2/8
0Vac
230Vac
Gnd
Summary of Contents for Compax 40CE
Page 138: ......