System Service Manual
−
SYSTEM COMPAX 40E, CE Marked
GE HEALTHCARE
with MPH 50/65/80 (including XT, RS 85 and Tomo
−
Link)
REV 9
DIRECTION 2154260
−
100
106
ILLUSTRATION 7
−
2
GROUNDING INTERCONNECTING MAP (SYSTEM MPH 50/65/80 GENERATOR WITH COMPAX 40E
−
MANUAL / SG100 / XT SUSPENSION)
SIGNAL POINT GND
PROTECTIVE
EARTH TERMINAL
REFERENCE TERMINAL
EQUIPOTENTIAL TERMINAL
MIS RUN − FIELD
SUPPLIED WIRE
SAFETY GROUND WIRE
SIGNAL GROUND WIRE
SAFETY & SIGNAL
GROUND WIRE
MIS RUN − CABLE
REFER TO OPTION
MAPS FOR MIS NUMBER
FACTORY INSTALLED WIRE
C
B
A
S
CHASSIS
000
LEGENDE
000
MAIN GROUND
TB1−GND
I/O PANEL GND
CARRIAGE
GND STUD
XT SUSPENSION
COMPAX 40E TABLE
2140362 ( TABLE Ground Cable )
2138905 ( Power Supply Cable )
46−216211G10
XT Ground Cable
( Supplied with XT )
2140362 ( SG Ground Cable )
SG 100 WALL STAND
MPH A7
Control Console
MPH A9
Printer−X
MPH A8
Program−X
I +
I −
MPH A3 A2
Main Power
MPH A3 A3
Inverter Board
MPH A3
A1
DC Power
MPH
A1
MPH A2
HV TANK
MPH A5
Real Electronic Panel
MPH 50/65/80 CABINET
MPH A1 A2
Stator Connector
Panel
MPH A6 A4
Wall Bucky
MPH A6 A5
Table Tomo
Collimator
MPH A6 A3
Distribution
Board
MPH A6 A6
II Sensor Board
MPH A6 A8
Plugs Panel
MPH A3
FL1
MPH A4
Front Electronic Panel
GND STUD
Ground Plug
GND STUD
5.1
2105836
Printer Ground Cable
(Supplied with Printer)
45553701
(Console Ground Cable)
Note:
All grounding cables are yellow−green.
27301
10708
27300
OPTIONAL
4.1
3.1
2.1
6.1
1.1
27235
8.1
46−316922G2
XT Ground Cable ( Supplied with XT )
REFER TO JOBCARD IST002
IN APPROPRIATE VERSION
MPH SERVICE MANUALS
FOR GROUND PLUG LAYOUT
MPH V1 SM:
2137354
−
100
MPH V2/V3 SM: 2165156
−
100
Summary of Contents for Compax 40CE