SECTION 4
ULTRANET COLLIMATOR INSTALLATION CHECK LIST
Complete the following checklist in Table 6
1 to ensure proper installation. If all of the answers to the
checklist questions are ‘yes’, then the collimator is properly installed.
TABLE 6
−
1
ULTRANET COLLIMATOR INSTALLATION CHECKLIST
CHECKLIST
YES
NO
1
Try to insert the collimator spanner wrench between the lock ring and the collimator mounting plate. Refer
to Illustrations 6
4. Does the space between the ring and the mounting plate prevent the wrench from being
inserted? (indicates full thread engagement)
2
Is the collimator roughly parallel to the X
−
ray tube? (If not, it could indicate a
potential cross threaded condition)
3
Are the threads of the mounting plate on the X
−
ray tube covered up by the locking ring or not visible? (indi-
cates full thread engagement)
4
Has the proper torque been applied to the collimator locking ring?
5
Have the 2 set screws been installed with Loctite?
6
Has the proper torque been applied to the 2 set screws?
ILLUSTRATION 6
−
4
CHECKING THE LOCKING RING AND COLLIMATOR MOUNTING PLATE SPACING