9-2
Manufacturer’s Guidance and Declaration Regarding
Electromagnetic Emissions
The BiliSoft 2.0 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BiliSoft 2.0 should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment Guidance
RF emissions CISPR 11
Class B
The BiliSoft 2.0 uses RF energy only for its internal function. Its
RF emissions are very low and are not likely to cause any
interference in nearby equipment.
Harmonic emissions IEC
61000-3-2
Class A
The BiliSoft 2.0 is suitable for use in all hospitals and domestic
residences directly connected to a public low voltage power
utility.
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Complies
Manufacturer’s Guidance and Declaration Regarding
Electromagnetic Immunity
The BiliSoft 2.0 is intended for use in the electromagnetic environment specified below. The customer
or the user of the BiliSoft 2.0 should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic
Environment
Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
+/- 8kV contact
+/- 15kV air
+/- 8kV contact
+/- 15kV air
Floors should be wood,
concrete or tile. If
floors are carpeted, the
relative humidity
should be at least 30%.
Electrical fast
transient/burst IEC 61000-
4-4
+/- 2kV for power supply
line.
100 kHz Repetition Rate
+/- 2kV for power supply
line.
100 kHz Repetition Rate
Electric power supplied
should be that of a
typical public low
voltage power utility.
Surge IEC 61000-4-5
+/- 1kV line to line
+/- 1kV line to line
Electric power supplied
should be that of a
typical public low
voltage power utility.
Voltage dips, shorts
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5 % Ut (>95 % dip in Ut)
for 0.5 cycle
At phase angles of :
<5 % Ut (>95 % dip in Ut)
for 0.5 cycle
At phase angles of :
Electric power supplied
should be that of a
typical public low
voltage power utility
1
.
Summary of Contents for BiliSoft 2.0
Page 1: ...BiliSoft 2 0 Phototherapy System Lighting by LumitexTM Operation Manual ...
Page 8: ...viii This page intentionally left blank ...
Page 12: ...xii This page intentionally left blank ...
Page 14: ...xiv This page intentionally left blank ...
Page 18: ...xviii This page intentionally left blank ...
Page 24: ...1 6 This page intentionally left blank ...
Page 32: ...2 8 This page intentionally left blank ...
Page 36: ...3 4 This page intentionally left blank ...
Page 43: ...5 1 Chapter 5 This chapter has been intentionally left blank ...
Page 44: ...5 2 This page intentionally left blank ...
Page 46: ...6 2 This page intentionally left blank ...
Page 56: ...9 6 This page intentionally left blank ...
Page 58: ...A1 2 This page intentionally left blank ...