9-4
Manufacturer’s Guidance and Declaration Regarding
Electromagnetic Immunity to RF Wireless Communication Devices
The BiliSoft 2.0 is intended for use in the RF Wireless environment specified below. The customer or
the user of the BiliSoft 2.0 should assure that it is used in such an environment.
Test
Frequency
(MHz)
Band
(MHz)
Service
Modulation
Maximum
Power
(W)
Distance
(m)
Immunity
Test Level
(V/m)
385
380-390
TETRA 400
Pulse
Modulation
18 Hz
1.8
0.3
27
450
430-470
GMRS 460
FRS 460
FM
+/- 5 kHz
deviation
1 kHz sine
2
0.3
28
710
704-787
LTE Band
13,17
Pulse
Modulation
217 Hz
0.2
0.3
9
745
780
810
800-960
GSM 800/900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse
Modulation
18 Hz
2
0.3
28
870
930
1720
1700-1990
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band
1,3,4,25 UMTS
Pulse
Modulation
217 Hz
2
0.3
28
1845
1970
2450
2400-2570
Bluetooth
WLAN
802.11 b/g/n
RFID 2450
LTE Band 7
Pulse
Modulation
217 Hz
2
0.3
28
5240
5100-5800
WLAN 802.11
a/n
Pulse
Modulation
217 Hz
0.2
0.3
9
5500
5785
Summary of Contents for BiliSoft 2.0
Page 1: ...BiliSoft 2 0 Phototherapy System Lighting by LumitexTM Operation Manual ...
Page 8: ...viii This page intentionally left blank ...
Page 12: ...xii This page intentionally left blank ...
Page 14: ...xiv This page intentionally left blank ...
Page 18: ...xviii This page intentionally left blank ...
Page 24: ...1 6 This page intentionally left blank ...
Page 32: ...2 8 This page intentionally left blank ...
Page 36: ...3 4 This page intentionally left blank ...
Page 43: ...5 1 Chapter 5 This chapter has been intentionally left blank ...
Page 44: ...5 2 This page intentionally left blank ...
Page 46: ...6 2 This page intentionally left blank ...
Page 56: ...9 6 This page intentionally left blank ...
Page 58: ...A1 2 This page intentionally left blank ...