1-2
Component/Control
Function
1.
Standby Switch – Turns the
unit ON/OFF. The green light in
the switch indicates that the
standby switch is turned on
and the unit has power.
2.
Hour Meter –The non-
resettable hour meter shows
the number of hours a light pad
has been active during the life
of the light box. If the Light Pad
connector is not fully inserted
into the light box, the LED is
automatically shut off and the
hour meter does not run. Note:
The hour meter is provided to
track LED life and is not
intended to be used to measure
therapy durations.
3.
Unit Overheated Indicator –
When this red indicator light is
ON, the unit has overheated.
See Troubleshooting Guide for
more details.
4.
LED Module Failure Indicator -
When the red indicator light
FLASHES, the LED module has
failed. See the Troubleshooting
Guide for more details.
5.
Light Pad Connector Port –
Where the light pad assembly
connects to the main unit. A
shutoff switch inside the port
automatically turns off the LED
module whenever the fiberoptic
light pad assembly is removed.
6.
Exhaust Air Vents – For proper
cooling it is important to keep
the air vents clear of
obstructions.
1
2
3
4
5
6
1
2
3
4
5
6
Summary of Contents for BiliSoft 2.0
Page 1: ...BiliSoft 2 0 Phototherapy System Lighting by LumitexTM Operation Manual ...
Page 8: ...viii This page intentionally left blank ...
Page 12: ...xii This page intentionally left blank ...
Page 14: ...xiv This page intentionally left blank ...
Page 18: ...xviii This page intentionally left blank ...
Page 24: ...1 6 This page intentionally left blank ...
Page 32: ...2 8 This page intentionally left blank ...
Page 36: ...3 4 This page intentionally left blank ...
Page 43: ...5 1 Chapter 5 This chapter has been intentionally left blank ...
Page 44: ...5 2 This page intentionally left blank ...
Page 46: ...6 2 This page intentionally left blank ...
Page 56: ...9 6 This page intentionally left blank ...
Page 58: ...A1 2 This page intentionally left blank ...