Revision C
ApexPro Telemetry Server
4-17
2001989-178
Maintenance: Checkout
To run disk check:
1.
Select Start > Shut Down > Log off, then click OK.
2.
Log back in as administrator, according to
“ApexPro Telemetry Server Service
3.
After the ApexPro Telemetry Server desktop is displayed, double-click My
Computer
4.
Right-click the disk drive you want to check and select Properties.
5.
Select the Tools tab.
6.
Under Error-checking, click Check Now.
7.
Check both selections and click Start.
8.
Click Yes in the Checking Local Disk window.
9.
Select Start > Shut Down >Restart, then click OK. The disk check is performed
as the ApexPro Telemetry Server reboots.
10. Repeat this procedure for the other drive.
Receiver Function/Available Receivers Check
Determine the Unity network devices that are communicating over the network by
using the Webmin Unity network function. See
“Unity Network Device Detection”
127(
This procedure requires a VNC connection. See
127(
Type help at any time to see a list of commands.3
1.
Select Start > Run
2.
Type cmd, then press Enter.
3.
A command window opens displaying the prompt:
C:\Documents and settings\cic>
4.
Type cd <space> ..\..\program files\marquette\pts\3.X (X= software version)
and press Enter.
5.
Verify the prompt:
C:\Program Files\marquette\PTS\X.X> (X= software version)
6.
In the command window at the PTS\X.X prompt, type ptsconfig and press
Enter.
7.
Type display receiver and press Enter.
8.
Note the number of unassigned receivers and subtract this number from the
number of admitted beds noted above.
Summary of Contents for ApexPro
Page 1: ...ApexPro Telemetry Server Service Manual 2001989 178 Revision C ...
Page 8: ...vi ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 9: ...Revision C ApexPro Telemetry Server 1 1 2001989 178 1 Introduction ...
Page 10: ...1 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 17: ...Revision C ApexPro Telemetry Server 2 1 2001989 178 2 Equipment Overview ...
Page 18: ...2 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 57: ...Revision C ApexPro Telemetry Server 3 1 2001989 178 3 Installation ...
Page 58: ...3 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 90: ...3 34 ApexPro Telemetry Server Revision C 2001989 178 Installation Completion For your notes ...
Page 91: ...Revision C ApexPro Telemetry Server 4 1 2001989 178 4 Maintenance ...
Page 92: ...4 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 111: ...Revision C ApexPro Telemetry Server 5 1 2001989 178 5 Troubleshooting ...
Page 112: ...5 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 135: ...Revision C ApexPro Telemetry Server 6 1 2001989 178 6 Parts Lists Drawings and Replacement ...
Page 136: ...6 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 191: ...Revision C ApexPro Telemetry Server A 1 2001989 178 A Technical Specifications ...
Page 192: ...A 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 198: ...B 2 ApexPro Telemetry Server Revision C 2001989 178 For your notes ...
Page 201: ......