Page 30 of 53
Copyright © 2007, General Electric Company
2023486-201 Rev C
WARNINGS
The following cautions must be observed to prevent problems during the rescue.
DANGER:
Fire and Explosion Hazard
Do not use in the presence of flammable gases (including concentrated oxygen) to avoid possible explosion or fire
hazard.
WARNING:
Shock Hazard
Defibrillation shock current flowing through unwanted pathways is potentially a serious electrical shock hazard. To
avoid this hazard during defibrillation abide by all of the following:
•
Do not touch the patient, unless performance of CPR is indicated
•
Do not touch metal objects in contact with the patient
•
Do not use in water
•
Keep defibrillation pads and ECG electrodes clear of other pads or metal parts in contact with
patient
•
Disconnect all non-defibrillator proof equipment from the patient before defibrillation
WARNING
: Shock and Possible Equipment Damage
Disconnect all non-defibrillator proof equipment from the patient before defibrillation to prevent electrical shock and
potential damage to the equipment.
CAUTION:
Use only Approved Equipment
Using batteries, pads, cables, or optional equipment other than those approved by General Electric may cause the
AED Pro to function improperly during a rescue.
CAUTION:
Possible Improper AED Pro Performance
Using pads that are damaged or expired may result in improper AED Pro performance.
CAUTION
: Possible Radio Frequency (RF) Susceptibility
RF susceptibility from cellular phones, CB radios and FM 2-way radio may cause incorrect rhythm recognition and
subsequent shock advisory. When attempting a rescue using the AED Pro, do not operate wireless
radiotelephones within 2 meters of the AED Pro – turn power OFF to the radiotelephone and other like equipment
near the incident.
CAUTION:
Possible Interference with Implanted Pacemaker
Therapy should not be delayed for patients with implanted pacemakers and a defibrillation attempt should be made
if the patient is unconscious and not breathing. The AED Pro has pacemaker detection and rejection, however with
some pacemakers the AED Pro may not advise a defibrillation shock.
Placing
Pads:
•
Do not place the pads directly over an implanted device.
•
Place the pad at least on inch from any implanted device.
CAUTION
: Moving the Patient During a Rescue
During a rescue attempt, excessive jostling or moving of the patient may cause the AED Pros to improperly
analyze the patient’s cardiac rhythm. Stop all motion or vibration before attempting a rescue.
Summary of Contents for AED Pro
Page 1: ...Operator s Manual 2023486 201rC...
Page 55: ......