Figure A1. STAR BIPHASIC WAVEFORM
Table A1 - Ultra-Low Current Responder AED Pro (all values are typical)
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25
1390 3.3 730 3.2
145-195
50
1420 4.5 915 3.2
130-175
75
1430 5.8 980 3.2
120-160
100
1435 7.0 1020 3.2 110-150
125
1440 8.3 1040 3.2 105-140
Table A2 – Low Variable Energy Waveform Responder AED Pro (all values
are typical
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25 1570 3.3 825 3.2
200-250
50 1600 4.5 1030 3.2 170-210
75 1610 5.8 1105 3.2 120-160
100 1615 7.0 1150 3.2 150-180
125 1620 8.3 1170 3.2 140-170
Table A3 – High Variable Energy Waveform Responder AED Pro (all values
are typical
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage
(Volts)
Duration
(ms)
Voltage
(Volts)
Duration
(ms)
Energy
(Joules)
25 1885 3.3 990 3.2
265-360
50 1920 4.5 1240 3.2 235-320
75 1930 5.8 1325 3.2 215-295
100 1940 7.0 1380 3.2 200-270
125 1945 8.3 1405 3.2 190-260