42
Cardiac Bypass
There are two types of cardiac bypass. Manual ventilation cardiac bypass is standard. VCV cardiac bypass is optional.
Manual ventilation cardiac bypass suspends alarms for patients on cardiac bypass when the ventilator is not
mechanically ventilating.
Systems with the VCV cardiac bypass option enabled can mechanically ventilate while in VCV mode. The VCV mode
is the only ventilation mode available while using VCV cardiac bypass.
Using Manual Ventilation Cardiac Bypass:
1.
Set the Bag/Vent switch to Bag.
2.
Select
Ventilation Mode
quick key.
3.
Select
Cardiac Bypass
, and then select
Start Cardiac Bypass
. The
Cardiac Bypass
message shows in the
waveforms and in the general message field when manual ventilation cardiac bypass is active.
Using VCV Cardiac Bypass:
1.
Start mechanical ventilation in VCV mode.
2.
Select
Ventilation Mode
quick key.
3.
Select
Cardiac Bypass
, and then select
Start Cardiac Bypass
.
•
PEEP is set to 5 cmH
2
0
•
TV settings of less than 170 ml prior to starting cardiac bypass remain at the set TV.
•
TV settings of more than 170 ml prior to starting cardiac bypass change to 170 ml.
•
The ‘VCV Cardiac Bypass’ message shows in the waveforms and in the general message field when VCV
cardiac bypass is active.