36
ecoFLOW menu components
1. Agent:
Shows Agent cost and flow information.
2. Agent Cost:
The cost of the current agent flow.
This value is determined by the agent flow multiplied
by the agent cost set by the Super User.
3. FiO
2
flow marker:
The graphical representation on
the flow tube of the FiO
2
flow value. This marker can
be removed by the Super User
4. Agent flow:
The measured value of the liquid agent
flow from the vaporizer. The agent flow may have a
delayed response.
5. FiO2 flow:
The minimum O
2
flow needed to
maintain the set inspired O
2
flow. This item can be
disabled by the Super User.
6. O
2
total (numeric):
The O
2
total equals the set of
O
2
flow. If N
2
O is the balance gas, this equals the set
O
2
flow. If Air is the balance gas, this is the set of O
2
flow plus 21% of the Air flow.
7. O
2
total (graphical):
The graphical representation
of the total O
2
flow.
8. N2O total and N2 total:
The N
2
O total equals the
set N
2
O flow. The N
2
total equals the N
2
in the Air
flow.
9. Flow bobbin:
The height of this represents the total
fresh gas flow delivered to the breathing system.
10. Total flow:
Shows Total Flow information.
I/min
ml/h
02 Total
Sev
€ 6.00 /h
Fi 35
Fi 35 02 flow
0.50
1
2
200
9
20
15
3
Total Flow
0.55
8
Agent
5
10
6
7
8
9
1
2
3
4