PSVPro™ Ventilation Mode
•
Pressure supported ventilation with apnea backup
•
Clinician sets the Psupport and PEEP levels, the patient establishes the rate, inspiratory flow and inspiratory time
•
Tidal volume is determined by the pressure, lung characteristics and patient effort
•
Apnea backup mode (SIMV PCV) is provided if the patient stops breathing
•
In backup mode, the Backup Mode active alarm is shown until PSVPro™ Ventilation Mode is reinstated or another
ventilation mode is selected
PSVPro mode settings:
•
Psupport
•
PEEP
•
Trig Window
•
Flow Trigger
•
End of Breath
•
Pmax
•
Backup Time
•
Pinsp
•
RR
•
Tinsp
•
Rise Rate
•
Exit Backup
Time
Pre
ss
ure
Peep
CPAP + PSV mode
•
Provides a constant support pressure for spontaneously breathing patients once the ventilator senses an
inspiratory effort
•
The patient establishes the rate, inspiratory flow, and inspiratory time
•
The tidal volume is determined by the pressure, lung characteristics, and patient effort.
•
Spontaneous breaths that occur are indicated by a color change in the waveform
CPAP + PSV settings:
•
Psupport
•
PEEP
•
Flow Trigger
•
End of Breath
•
Pmax
•
Pinsp
•
Minimum RR
•
Tinsp
•
Rise Rate
34
Time
Pre
ss
ure
Peep
T
insp
T
insp