SIMV PCV-VG mode
•
Delivers a set rate of pressure controlled breaths with a guaranteed volume
•
The ventilator will adjust the inspiratory pressure needed to deliver the set tidal volume breath-by-breath so that
the lowest pressure is used.
•
The patient’s compliance is determined from the volume controlled ventilation breath and the inspiratory
pressure level is then established for the next PCV-VG breath.
SIMV PCV-VG mode settings:
•
TV
•
RR
•
Flow Trigger
•
Psupport
•
PEEP
•
Tinsp
•
Pmax
•
Trig Window
•
End of Breath
•
Rise Rate
•
PSV Rise Rate
Tinsp
Texp
Trigger
window
Trigger
window
Fl
ow
TV
TV
Time
Pre
ss
ure
Time
Variable pressure to
deliver desired TV
Tinsp
Texp
Peep