41
Spirometry setup
Select
Setup Loops
from the Spirometry menu
to access the Setup loops menu.
1. Spirometry Scaling:
Set the scales of the
spirometry loop graph. The available
settings for the volume, Paw, and flow
graph axis are dependent on the set
patient type of adult or pediatric.
•
Auto:
automatically adjusts the
volume, Paw, and flow axis of the loop.
•
Linked:
links the adjustment of the
volume, Paw, and flow axis of the loop
graph together.
•
Indep.
allows the axis of the loop
graph to be changed separately for
the volume, Paw, and flow axis.
2. Patient and Sensor:
Select
Adult
or
Pedi
depending on the sensor used.
3. Data Source:
Select
Patient
to have
spirometry data sourced from the airway
module or
Vent
to have spirometry data
sourced from the ventilator.
4. Show MV or TV:
Set to
TV
to show TVinsp
and TVexp on the spirometry split screen.
Set to
MV
to show MVinsp and MVexp on
the spirometry split screen.
1
2
3
4