Care and Maintenance
10-24
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
Grounding continuity
Measure the resistance from the third pin of the attachment plug
to the exposed metal parts of the case. The ground wire
resistance should be less than
0.2
ohms. Reference the
procedure in the IEC60601-1.
Figure 10-2. Ground continuity test
DANGER
ELECTRIC SHOCK HAZARD. THE PATIENT MUST NOT BE
CONTACTED TO THE EQUIPMENT DURING THIS TEST.
1. GROUND
PIN
2. OHMMETER
3. LOGIQ C9 Series
4. ACCESSIBLE METAL PART:
• MONITOR HOUSING
• PEAR PANEL CONNECTOR
• ANY CASTER/WHEEL SUPPORT
Summary of Contents for 5489329
Page 5: ...LOGIQ C9 Series Basic Service Manual i 3 5482298 100 English Rev 5 ...
Page 6: ...i 4 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 7: ...LOGIQ C9 Series Basic Service Manual i 5 5482298 100 English Rev 5 ...
Page 8: ...i 6 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 9: ...LOGIQ C9 Series Basic Service Manual i 7 5482298 100 English Rev 5 ...
Page 10: ...i 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 11: ...LOGIQ C9 Series Basic Service Manual i 9 5482298 100 English Rev 5 ...
Page 12: ...i 10 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 13: ...LOGIQ C9 Series Basic Service Manual i 11 5482298 100 English Rev 5 ...
Page 18: ...i 16 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 52: ...Introduction 1 28 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 76: ...Site Preparations 2 24 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 118: ...System Setup 3 42 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 206: ...Renewal Parts 9 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 242: ...Care and Maintenance 10 36 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 247: ...GE ...